Tiny flowmeter

I. Overview
DXBTMF series micro-small metal tube float flowmeter is a variable area flow meter based on float position measurement. The new design makes it possible to measure small flows. Particularly suitable for the measurement of small flow rates of liquids and gases. Full metal structure with small volume, small pressure loss, and large range ratio (10: 1), widely used in various industries in complex and harsh environments, flow measurement and process for small flow rates, low flow rates, and various harsh media conditions Control, especially suitable for measuring turbid, opaque or corrosive fluids.
DXBTMF series micro-small metal tube float flowmeter, according to different user needs and different occasions, there are multiple measurement forms for users to choose; according to the output form, there are local instruction type and remote transmission output type; according to explosion-proof requirements, It can be divided into two types: ordinary type and intrinsically safe type.
The remote transmission part of DXBTMF series micro-small metal tube float flowmeter uses the international advanced non-contact and non-hysteresis detection magnetic field sensor to detect the change of the magnetic field angle. It is also equipped with a high-performance MCU. The host computer software sets parameters such as linear calibration of the flowmeter. This type of intelligent signal transmitter has high accuracy and reliability, high cost performance, and can completely replace imported instruments of the same type.
The design and manufacture of DXBTMF series micro-small metal tube float flowmeters also consider the user's process flow requirements. There are DXBTMF-V for bottom-in, top-out installation and DXBTMF-H for side-in, side-out installation.
Structure and principle

2. DXBTMF series micro-small metal tube float flowmeter consists of the following parts:
Ontology
2, measuring tube
3.Float
4.Header
5.Magnetic coupling system
6.Electrical cables

There are three types of wetted materials for the sensor: stainless steel, Hastelloy, and titanium. Users can choose different wetted materials according to different process pressures and corrosive requirements of the medium to meet the needs of process pressure resistance and medium corrosion protection.
The flow measurement is performed by the transmitter inside the indicator through coupling the magnetic steel to sense the change in the position of the float to complete the flow indication and the remote transmission of the signal. When the measured medium flows from the bottom to the top of the measuring tube, the float is balanced by the force of gravity, buoyancy, and fluid velocity on the float's vertical upward driving force. The float is relatively stationary at a certain position. The annulus area and fluid velocity change, and the position of the float corresponds to the flow rate of the measured medium.

Third, the characteristics of the smallest measurable flow range: 0.1 to 1 l / h water
Side in and out type with outlet needle valve or inlet needle valve can take two-wire 4-20mA analog output (24VDC power supply)
Optional matching constant flow valve installation distance is minimum 90mm (side in and side out)
All stainless steel structure using temperature range: -40-+ 80 ℃
Damping time setting range: 0-10 seconds. Settable accuracy: <± 2.5%, temperature drift: <± 0.1% / 10 ℃.
All configuration operations can be performed on the flowmeter through an isolated USB-UART converter and host computer software.
The main parameters can be set through the four keys on the flow meter and connected to an external LCD display.

4. Technical parameter measurement range Water (20 ℃) 1 ~ 3000 l / h
Air (0.1013MPa 20 ℃) 0.5 ~ 90000 l / h
Turndown ratio 10: 1
Accuracy class 2.5 / 4.0 working pressure PN16MPa
Medium temperature standard -40 ℃ ~ + 120 ℃
Ambient temperature -40 ℃ ~ + 80 ℃
Medium viscosity 1/4 ”NPT, 3/8” NPT 1/2 ”NPT≤5mPa.s

3/4 ”NPT, 1” NPT ≤250mPa.s

Output Standard signal: Two-wire system 4 ~ 20mA (24VDC power supply)
Power supply 24VDC ± 20%
Connection method Side-in test type: 1/4 ”NPT
Bottom-in, top-out type: 1/4 ”NPT, 3/8” NPT, 1/2 ”NPT, 3/4” NPT, 1 ”NPT female thread electrical interface M8 × 1 or customer-supplied specifications

Protection class IP65
Explosion-proof mark Intrinsically safe type: ExiaIICT3 ~ 6